CAUTION (CONTINUED)
•
The cot fastener system must be installed by qualified personnel. 1/4" steel backing plates (or equivalent) are
required beneath the mounting surface (ambulance floor or tray system) at all mounting points.
•
Always clear any obstacles that may interfere and cause injury to the operator or patient before operating the
product.
•
Do not apply the Steer-Lock and attempt to push the product sideways.
•
Do not entangle the restraint straps in the base frame when you raise or lower the cot.
•
Always secure the defibrillator platform to the product when you use the defibrillator platform.
•
Always use and adjust the straps that are provided with the defibrillator platform to secure the defibrillator.
•
Always change the installation location or adjust the straps for your specific defibrillator size or shape.
•
Do not load the defibrillator platform above the safe working load of 30 lb (13.6 kg).
•
Do not load the oxygen bottle holder above the safe working load of 15 lb (6.8 kg).
•
Do not use two oxygen bottle holders at the same time.
•
Do not steam clean or ultrasonically clean the product.
•
Do not exceed 180° F (82° C) as the maximum water temperature.
•
Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you are using a hand held wand to wash the
product, keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
•
Always allow to air dry.
•
Do not clean, disinfect, service, or perform maintenance while the product is in use.
•
Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive in nature
and may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue on
the surface of the product and may cause premature corrosion of critical components.
•
Always use authorized parts to avoid the risk of product damage.
www.stryker.com
Summary of safety precautions
6100-109-005 REV B