After having sanitised the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of
the methods described below.
method A: The disinfectable accessories are 1A-1B-1C-2-3-4-5-6.
The disinfectant must be an electrolytic chlorine oxidiser (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Directions:
- Fill a container of suitable size to hold all of the parts to be disinfected with a
solution of potable water and disinfectant, according to the proportions indicated
on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of
air bubbles on the parts. Leave the parts immersed for the amount of time indicated
on the packaging of the disinfectant associated with the concentration chosen for
the solution.
- Remove the disinfected parts and rinse with plenty of warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Sanitise the accessories 1A-1B-1C-2-3-4-5-6 by boiling them in water for 10
minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Sanitise the accessories 1A-1B-1C-2-3-4-5-6 with a hot steam steriliser
for baby-bottle (not the microwave type). Perform the process faithfully following
the instructions of the steriliser. To ensure that the disinfection is effective, choose a
steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on
a paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair
dryer).
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND
STERILISATION
Before disinfection or sterilisation, sanitise the nebuliser cup and the accessories,
choosing one of the methods described below.
method A: Sanitise accessories 1A-1B-1C-2-3-4-5-6 under potable hot water
(approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories 1A-1B-1C-2-3-4-5-6 in the dishwasher with a hot cycle.
DISINFECTION
The disinfectable accessories are 1A-1B-1C-2-3-4-5-6.
The disinfectant must be an electrolytic chlorine oxidiser (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Directions:
9