• Do not immerse the device in water; in the event of immersion disconnect the plug immediately. Do not remove
or touch the immersed device; unplug the power cable first. Immediately bring the device to an authorised FLAEM
service centre or to your trusted dealer.
• Do not wash the device under running water or by immersion and keep it safe from being sprayed by water or other
liquids.
• Do not expose the device to particularly extreme temperatures.
• Keep the device and power supply away from sources of heat, direct sunlight or excessively hot places.
• Repairs may only be performed by authorised FLAEM personnel following the information in the restricted area of
the website www.flaem.it. Unauthorised repairs void the warranty and may pose a danger for the user.
WARNING: Do not modify this device without authorisation from the manufacturer.
• The Manufacturer, the Dealer and the Importer shall be held responsible for safety, reliability and performance only
if: a) the device is used in compliance with the instructions for use b) the wiring where the device is being used is
in compliance with safety regulations and current laws.
• The Manufacturer must be contacted about any problems and/or unexpected events concerning operation and if
any clarifications on use, maintenance or cleaning are needed.
• Should your device fail to provide the expected performance, contact the authorised Flaem service centre for clar-
ifications.
• Ensure the connections and the canister lid have been carefully sealed in order to prevent loss of suction.
• Do not overturn the canister while connected to the operating device as the liquid may be aspirated into the
device, thus damaging the pump. Should this occur, immediately switch off the suction machine, then empty and
clean the canister (send it to an authorised FLAEM service centre).
• When the safety device (10) is triggered, suction is halted. Switch off the device, empty the canister (9) and carry out
the cleaning procedure.
• We recommend personal use of the accessories, the collection canisters and the connecting tubing to prevent risk
of contagious infection.
• The manual vacuum flow control (15) is a sterile disposable product and must be replaced after each use.
• Check the expiry date on the original package of the manual vacuum flow control (15) and ensure the sterile pack-
age is undamaged. Should the package not be intact contact the authorised Flaem service centre or your trusted
dealer. FLAEM NUOVA disclaims any liability for harm to the patient related to deterioration of the aforementioned
sterile package due to third parties handling the original packaging of the entire device.
• The 12V DC power cable for vehicle power outlets is equipped with a safety fuse, which can be inspected in the
event of a fault.
• The power supply provided has been specifically designed for use with the ASPIRA Go series aspirators. Never use
the power supply with other devices or for uses other than that specified in this manual and never use the ASPIRA
Go series aspirators with other power supplies.
• If the device is not used for a long period (6 months), it is advisable to remove the battery before storage.
• The hydrophobic/antiviral and antibacterial filter supplied with Flaem aspirators must be strictly replaced
for each new patient, or if the said filter is saturated. Should the aspirator be used directly on the patient,
the filter for Single-patient use must be replaced within a maximum period of TWO MONTHS, or if the said
filter is saturated.
INSTRUCTIONS FOR USE
Before each use, the accessories must be carefully inspected to ensure the absence of dust, incrustation, clots or liquid
substances both inside the connecting tubing and in the canister and its respective lid. Furthermore, they must be
cleaned following the instructions rigorously as stated in the "CLEANING, SANITISATION, DISINFECTION, STERILISA-
TION" paragraph. We recommend personal use of the accessories, the collection canisters and the connecting tubing
to prevent risk of contagious infection.
1. Operation with internal battery:
1.1. The device is supplied with the battery partially charged and it is recommended to charge it before use. Insert
the battery as shown on page 1. Release, lift and remove the lid of the battery compartment located at the back
of the device (dwg.1,2,3); insert the battery, fitting the connector properly (dwg.4,4a); put the battery compart-
ment lid back in place and close it (dwg.5,6). Recharge the battery by following the instructions in section 2.3.1.
1.2. Switch the device on by pressing the switch (2).
1.3. At the end of each suction procedure, always turn the switch (2) to 0 (even if the battery is flat) to save the battery
life. If during use the red LED (1c) flashes and the buzzer sounds, it means that the device's battery charge is
about to run out, in this case recharge the device as soon as possible.
CAUTION: Periodically check the status of the battery at least once a month to avoid the need to use the device in
an emergency and not to support the internal power source. It is recommended to charge the battery at least once
every 3 months .
2.
Operation and charging with the vehicle power outlet (12V DC), or with the multi-voltage power supply:
12