Manual Cleaning And Disinfection - Erbe 20190-148 Manual Del Usuario

Mango de nervtest
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
OBJ_DOKU-182711-004.fm Seite 22 Dienstag, 22. November 2016 3:25 15
5.8

Manual cleaning and disinfection

Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
1. Prepare a cleaning bath and a separate disinfecting bath ac-
cording to the manufacturer's instructions.
2. Immerse the product completely in the cleaning bath, with-
out the product touching other parts in the bath. Observe the
recommended immersion time.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Rinse the surfaces with sterile, deionized water (at least
1 minute).
5. Repeat the preceding cleaning steps if the product still
shows visible contamination.
6. Immerse the product completely in the disinfectant bath,
without the product touching other parts in the bath. Ob-
serve the recommended immersion time.
7. Rinse the surfaces with sterile, deionized water (at least
1 minute).
8. Dry the product with filtered compressed air.
22
Tabla de contenido
loading

Tabla de contenido