The  device  is  delivered  with  default  setting  activated  beeper.  
 
3.    
Before  each  measurement  
3.1     Selecting  the  correct  cuff  
Riester  offers  different  cuff  sizes.  Select  the  cuff  size  to  match  the  circumference  of  your  patient's  
upper  arm  (measured  by  close  fitting  in  the  centre  of  the  upper  arm).  
 
 
Cuff  size  
S  (option)  
M  
L-­XL  
 
3.1.1  
Always  ensure  that  the  correct  cuff  size  is  used  (size  markings  on  the  cuff)  
3.1.2  
Contact  your  local  Riester  service  if  the  enclosed  cuffs  (12)  do  not  fit.  
3.1.3  
Connect  the  cuff  to  the  device  by  clicking  the  cuff  connector  (13)  on  to  the  cuff  socket  (8).  
 
3.1.1 Always ensure that the correct cuff size is used (size markings on the cuff)
3.1.2 Contact your local Riester service if the enclosed cuffs (12) do not fit.
Only  use  Riester  cuffs.  
3.1.3 Connect the cuff to the device by clicking the cuff connector (13) on to the cuff socket (8).
3.2     Fitting  the  cuff  
3.2.1  
Remove  close-­fitting  garments  from  the  upper  arm  of  the  patient.  To  avoid  constriction,  shirt  
Only use Riester cuffs.
 
sleeves  should  not  be  rolled  up  -­  they  do  not  interfere  with  the  cuff  if  they  are  laid  flat.  
3.2.2  
Position  the  cuff  on  the  upper  arm  (right  or  left),  so  that  the  tube  points  in  the  direction  of  the  
 
lower  arm.  
3.2.3  
The  artery  mark  on  the  cuff  must  lie  over  the  artery  which  runs  down  the  inner  side  of  the  arm.  
3.2.4  
Make  sure  that  the  cuff  is  positioned  2-­3  cm  above  the  elbow.  

3.2 FITTING THE CUFF

3.2.5  
Secure  the  cuff  with  the  Velcro  and  make  sure  the  cuff  is  comfortably  attached  and  not  too  
 
tight.    
3.2.6  
Lay  the  arm  of  the  patient  on  the  table  (palm  upwards),  so  that  the  cuff  is  at  the  same  height  
3.2.1 Remove close-fitting garments from the upper arm of the patient. To avoid constriction,
 
as  the  heart.  
3.2.7  
Make  sure  that  the  tube  is  not  kinked.  
shirt sleeves should not be rolled up - they do not interfere with the cuff if they are laid
flat.
3.3     Selecting  the  measurement  mode  
3.2.2 Position the cuff on the upper arm (right or left), so that the tube points in the direction
This  device  allows  you  to  measure  in  three  different  measurement  modes.  Select  between  Standard  
of the lower arm.
BP  Measurement  1x  (30),  Triple  BP  Measurement  3x  (31)  and  Manual  BP  Measurement  (32)  by  
pressing  repeatedly  the  Mode  button  (2).  The  current  set  measurement  mode  is  on  the  display  shown  
3.2.3 The artery mark on the cuff must lie over the artery which runs down the inner side of
by  the  related  symbol.  
the arm.
 
3.2.4 Make sure that the cuff is positioned 2-3 cm above the elbow.
4.    
Performing  BP  measurements  in  different  measurement  modes  
3.2.5 Secure the cuff with the Velcro and make sure the cuff is comfortably attached and not
4.1     Standard  blood  pressure  measurement  (1x)  
too tight.
4.1.1  
Select  the  standard  BP  measurement  mode  by  pressing  repeatedly  the  Mode  button  (2)  until  
 
1x  (30)  appears  on  the  display.    
3.2.6 Lay the arm of the patient on the table (palm upwards), so that the cuff is at the same
4.1.2  
Press  the  Start/Stop  button  (3)  to  start  the  measurement.  
height as the heart.
4.1.3  
The  cuff  will  now  pump  up  automatically.  The  inflation  is  indicated  by  the  blinking  cuff  inflation  
 
3.2.7 Make sure that the tube is not kinked.
symbol  (28).  The  patient  should  relax,  should  not  move  and  should  not  tense  his  arm  muscles  
 
until  the  measurement  result  is  displayed.  He  should  breathe  normally  and  not  talk.  
4.1.4  
When  the  correct  pressure  is  reached,  the  pumping  stops  and  the  pressure  falls  gradually.  
 
The  deflation  is  indicated  by  the  blinking  cuff  deflation  symbol  (27).  If  the  required  pressure  
 
was  not  reached,  the  device  will  automatically  pump  some  more  air  into  the  cuff.  

3.3 SELECTING THE MEASUREMENT MODE

4.1.5  
During  the  measurement,  the  pulse  indicator  (25)  flashes  in  the  display  and  a  beep  sound  
 
appears  each  time  a  heartbeat  is  detected.  
4.1.6  
The  result,  comprising  the  systolic  (17)  and  the  diastolic  (18)  blood  pressure,  pulse  rate  (19)  
 
and  mean  arterial  pressure  (MAP)  (22).  The  results  of  pulse  rate  and  mean  arterial  pressure  
This device allows you to measure in three different measurement modes. Select between
 
alternating  appear  on  the  display  every  2  seconds.    
Standard BP Measurement 1x (30), Triple BP Measurement 3x (31) and Manual BP Measurement
4.1.7  
When  the  device  has  finished  measuring,  remove  the  cuff.  
(32) by pressing repeatedly the Mode button (2). The current set measurement mode is on the
4.1.8  
Switch  off  the  device  (The  monitor  does  switch  off  automatically  after  approx.  1  min.).  
display shown by the related symbol.
For  circumference  of  upper  
arm  
14-­22  cm  (5.5  –  8.7  inches)  
22  -­  32  cm  (8.7  -­  12.6  
inches)  
32  –  52  cm  (12.6  -­  20.5  
inches)  