WARNING: If you see any body fluid in the connector opening, you must discard
the transmitter. Because the transmitter contains a battery, do not
discard in a bio-waste container. Instead, continue to clean and
disinfect the transmitter, and then discard according to local
regulations for battery disposal (non-incineration).
 3
Attach the tester to the transmitter to create a waterproof seal.
 4
If there is adhesive residue on the transmitter, see Removing adhesive residue, on
page 17.
 5
Rinse the transmitter under room temperature tap water for at least one minute and
until visibly clean. Make sure all hard-to-reach areas are rinsed completely.