INSTRUCTIONS FOR USE

4.3 Sterilization Instruction:

Sterilization directions apply to the S3 Arm Sleeve Connector (AR-1650-02) and the
S3 Lateral Traction Sling Connector (AR-1650-03).
AR-1650-02 Arm Sleeve Connector
Warning: Failure to properly sterilize device could cause exposure to
contaminant.
Steam Sterilization for AR-1650-02 & AR-1650-03
Sterilizer Type:
Preconditioning Pulses:
Minimum Temperature:
Full Cycle Time:
Minimum Dry Time:
Cool-Down Time:
Sterilizer Type:
Minimum Temperature:
Full Cycle Time:
Minimum Dry Time:
Cool-Down Time:
Document Number: D-720764
Version: C
AR-1650-03 Lateral Traction Sleeve Connector
Prevacuum
4
132°C
4 minutes
30 minutes
30 Minutes outside of chamber on a wire rack
Gravity
132°C
15 Minutes
50 Minutes
30 minutes outside of chamber on a wire rack