Hoefer HE-PLUS Manual Del Usuario página 19

Tabla de contenido
CAUTION: Do not jar or bump the 
gel box once the lid is place. 
The electrical connection is 
made by gravity once the lid is in 
position. While this design helps 
to minimize sample disturbance 
during lid placement, it also may 
result in a disruption of power 
to the unit if the lid or unit are 
disturbed during the run. 
CAUTION: DO NOT EXCEED THE 
MAXIMUM OPERATING VOLTAGE 
OF 150 VOLTS. 
G. To Pause a Run and/or Change Parameters 
1
To pause the run select Pause. During the pause mode 
the voltage amperage or time can be changed by 
highlighting the function and using the arrow keys then 
pressing enter. Once the changes have been made the 
start button can be pressed to resume the run.
2
To stop the run press the pause button for 30 seconds. 
Stop will appear.
H. Sample Electrophoresis 
1
The maximum suggested applied voltage for the 
electrophoresis of DNA in agarose gels using the 
Hoefer HE-PLUS Gel System is 150 volts. 
2
In a 1% TBE gel, this translates into a run time of 
approximately 1 hour. Lower voltages may be used, 
of course, and as a general rule, a 70 V run will take 
twice as long as a 145 V run. Higher voltages may 
be used to decrease run time, however, if the unit is 
being operated at higher voltages than 150 V, the heat 
generated during electrophoresis may decrease sample 
resolution. Such artifacts may be avoided by running 
the unit in a cold room or adding 1X electrophoresis 
buffer "ice cubes" to keep the unit properly cooled.
3
Follow the sample migration into the gel using the 
loading dye as an indicator. (See page 12 for the 
Sample Loading Buffer recipe.) Allow the samples to 
migrate until the fragments have separated, normally 
until the bromophenol blue dye front has migrated ¾ 
of the way down the gel.
 p9
Tabla de contenido
loading

Tabla de contenido