8.5 Dry
After cleaning and disinfection, please dry, it is recommended to use compressed air to dry.
134°C
8.6 Sterilization
The sterilization temperature recommended by the manufacturer is 134 °C.
Autoclave for 18 minutes at 134 °C.
1) Scrub dirt and debris from the contra-angle, and wipe clean with alcohol-immersed cotton swab or cloth. Do not
use a wire brush.
2) Insert into an autoclave pouch and seal the pouch.
3) Keep the contra-angle in the autoclave pouch to keep it clean until you use it.
CAUTION
– Do not was soak or wipe o the contra-angle within the oxidation potential solution (strong acid/super acid
solution) or sterilized solution.
– Do not autoclave the equipment with chemical residual: it may cause a peel of the plating or exert a harmful
infl uence on internal parts.
– Store it in a place where the handpiece is not subject to air with dust, sulfur or salinity.
– Do not touch the contra-angle just after the autoclave.
26

9. Transport and Storage

Operating Temperatures: +5 - +40 °C
Storage Temperatures: -10 - +55 °C
Operating Humidity: ≤ 80 %RH
Storage Humidity: ≤ 93 %RH

10. Standard Symbols

Attention
Certifi ed to MDD93/42/EEC
134°C
Autoclave at 134 °C
Manufacturer
European Union agent
EC REP
Thermo-Disinfector
27