Measuring ranges:
500 V : 10 k ... 1.999 T
1000 V : 10 k ... 3.999 T
2500 V : 10 k ... 9.99 T
5000 V : 10 k ... 9.99 T
Var 50 V ... 5000 V: to be interpolated between the fi xed values above.
Accuracy
Test voltage
Specifi ed
1.000...3.999 M
measurement
range
Resolution
Accuracy
Test voltage
Specifi ed
400...999 M
measurement
1.000...3.999 G
range
1 M
Resolution
Accuracy
±5% + 3 pt
Accuracy in variable mode
To be interpolated between the values of the table above and per § 7.2.2.
Measurement of DC voltage during insulation test
Specifi ed measurement range
Resolution
Accuracy
Measurement of DC voltage during discharge phase of an insulation test
Specifi ed measurement range
Resolution
Accuracy
500 V - 1000 V - 2500 V - 5000 V
4.00...39.99 M
10...999 k
1 k
500 V - 1000 V - 2500 V - 5000 V
4.00...39.99
40.0...399.9
G
G
10 M
100 M
40.0...99.9 V
0.1 V
25...5100 V
0.2% Un
5% + 3 pt
72
40.0...399.9 M
10 k
±5% + 3 pt
1000 V - 2500 V
5000 V
400...999 G
2.000...
1.000...1.999 T
3.999 T
1 G
±15% + 10 pt
100...1500 V
1 V
1% + 1 pt
100 k
2500 V
5000 V
4.00...
9.99 T
10 G
1501...5100 V
2 V