Sartorius YAS02CI Instrucciones De Instalación página 4

Idiomas disponibles
  • ES

Idiomas disponibles

Power
100V-240V
50/60Hz
15VA
Power
15.5V-24V =
13V-17V ~ 50/60 Hz
12W
4
Loosen the 2 clamp screws (1) and disconnect the
2 power cable leads from the power module.
Disconnect the ground wire (earth) (2).
Unplug the cable that connects the power
module to the main PCB (3).
Open the cable gland and remove the power cord
from the gland.
Remove the fastening screws from the power
module (4).
Remove the power module from the indicator.
Remove the input voltage label ("100V–240V")
from the back of the indicator housing.
Affix the enclosed voltage label in place of the
old label.
loading