WARNING: Rectal Light Handle and/or Power Supply is not suitable for
use in the presence of a flammable anesthetic mixture with air or
oxygen or nitrous oxide. An explosion may result.
WARNING: Lamp generates heat and should remain off until just prior
to and immediately after use. Do not place Rectal Light Handle onto a
flammable surface if lamp is illuminated or hot.
WARNING: Disconnect power supply from outlet prior to any
maintenance.
WARNING: Only use the 6V power supplies described within this
manual with the 73210 Rectal Light Handle. Failure to observe this
warning may cause damage to the power supply and/or non-approved
device
WARNING: The use of accessories, transducers, and/or cables other
than those specified, with the exception of those sold by the
manufacturer as replacement parts for internal components, may
result in increased emissions or decreased immunity of the equipment
or system.
CAUTION: U.S. Federal Law restricts this device to sale by or on the
order of a physician.
CAUTION: Rectal Light Handle Illumination is intense.
CAUTION: This endoscopic accessory is for use with the Welch Allyn
devices listed in the table on page 1-6 only. Do not use the illuminator
for any other use.
CAUTION: Use only the cleaning methods described in this manual.
CAUTION: There are no user-serviceable parts inside the device except
the lamp. Contact the authorized Service Center listed on page 1-20 for
service related issues.
CAUTION: The 73210 Rectal Light Handle is designed for connection
only with the power source and appropriate country voltage and outlet
style listed within this manual No other use of these products is
intended or advised.
CAUTION: Medical Electrical Equipment needs special precautions
regarding EMC and needs to be installed and put into service according
to the EMC information starting on page 1-12.
CAUTION: Portable and Mobile RF Communications Equipment can
affect Medical Electrical Equipment.
1-2