Maintenance
Preventative Maintenance
Examine cords, connections, and transformer case for wear or damage
monthly.
After every use, clean the Rectal Light Handle according to the Cleaning
Section of this manual.
Lamp Replacement
O-ring
NOTE: The lamp is the only user-serviceable part on the Rectal Light
Handle (REF 73210). Refer to the above figure as a visual aid to replace
the lamp.
1.
Disconnect the system power supply from the wall outlet and then
disconnect the Rectal Light Handle from the power supply cord. Allow lamp
to cool for a minimum of 5 minutes.
2.
Unscrew the Rectal Light Handle outer sleeve counter clockwise (CCW) to
expose the lamp as shown.
3.
Unscrew the lamp CCW from the Rectal Light Handle body.
4.
Inspect the lamp socket of the Rectal Light Handle for damage or
irregularities. Contact the authorized Service Center listed on page 1-20 if any
irregularities are found.
5.
Using a new Welch Allyn lamp, verify the lamp number is REF 07800 and
that the o-ring is present and in good condition.
6.
Screw the lamp clockwise (CW) into the Rectal Light Handle body and
tighten sufficiently to compress the lamp o-ring.
7.
Screw the Rectal Light Handle sleeve CW onto the Rectal Light Handle body
and tighten sufficiently to secure to the handle.
Lamp REF 07800
1-7
CCW
CW
Rectal Light Handle