Work On The Electrical Equipment Of The Device; Personal Protective Equipment; Safety Instructions Concerning Operation Of The Device; Installation - Sartorius PMA.Evolution Instrucciones De Instalación

Ocultar thumbs Ver también para PMA.Evolution:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Only operate the device when it is safe and in perfect
t
working order.
Immediately disconnect the damaged device from the
t
power.
Have any malfunctions or damage repaired immediately
t
by Sartorius Service personnel.
2.6 Work on the Electrical Equipment of the
Device
Work on and modifications to the electrical equipment of the
device may only be carried out by Sartorius Service personnel.
The device may only be opened by Sartorius Service personnel.
Seal on Scales Verified for Use in Legal Metrology
Legislation requires that a seal be affixed to
verified scales. On Sartorius devices, this seal takes
the form of a sticker with the "Sartorius" logo. If
the seal is removed, the validity of verification will
become void and you must have your scale
re-verified. For verified scales for use in the EEA,
the declaration of conformity set out in the
calibration and supplied here shall apply. Please
keep it in a safe place.

2.7 Personal Protective Equipment

Personal protective equipment protects against risks arising
from the material being processed.
When the workplace or the process in which the device
t
is used requires personal protective equipment: Wear
personal protective equipment.
2.8 Safety Instructions Concerning Operation
of the Device
− Take care that the glass panel of the operating display is not
damaged (e.g., by falling objects, impact, or extreme
pressure). If the glass panel is damaged, disconnect the device
from the power supply immediately.
− The surface of the operating display should not be touched
with pointed, sharp, hard, or rough objects. You should only
use the touch pen provided or your fingertips. Do not use
parts of clothing (e.g., sleeves) or sponges for cleaning
because these can scratch the surface (e.g., due to rivets or
buttons in the sleeve or sand in the sponge).
− Avoid generating static electricity on the glass panel of the
operating display and plastic casing.
Danger of Damage to the Scale!
Never close a paint can using a hammer while
it is still on the weighing pan.
When closing, place the paint can on a firm,
stable surface.
3

Installation

3.1 Scope of Delivery

Item
Large weighing pan: d 233 mm x
Small weighing pan: d 180 mm -
USB cable
AC adapter YEPS01-USB
Installation instructions

3.2 Unpacking

Procedure
Open the packaging, making sure to remove all parts
t
carefully.
After unpacking the device, check it immediately for any
t
external damage.
If the device is stored temporarily: Store the device
t
according to the ambient conditions (ambient conditions
see Chapter "8.1 General Data," page 10).
Save all parts of the original packaging for any future
t
transport. All cables should be unplugged when
transporting.

3.3 Selecting a Setup Location

Select the right setup location:
− Set up the device on a stable, even surface that is not
exposed to vibrations.
− Maintain free access to the device at all times.
Choose a location that is not subject to the following negative
influences:
− Heat (heater or direct sunlight)
− Drafts from open windows, AC systems, and doors
− Extreme vibrations during weighing
− Heavy "traffic areas" (personnel)
− Extremely high humidity
− Electromagnetic fields
− Extremely dry air
Acclimatization
Condensation from humidity can form on the surfaces of a
cold device when it is brought into a warm area. You should
therefore let a device that has been disconnected from its
power source acclimatize for approximately 2 hours before
reconnecting it to the supply voltage.
Installation Instructions EVO1Y | LAB1Y
English
PMA.Evolution PMA.HD
-
x
x
x
Optional
Optional
x
x
5