Electrical Connection; Wiring Details - SICK PBS Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Operating Instructions

Electrical connection

Connect the instrument to earth via the pressure connection.
For power supply, use a circuit with energy limitation (EN/UL/IEC 61010-1, section 9.3)
with the following maximum values for the current: with L + = 35 V (DC): 4.0 A. Provide a
separate switch for the external power supply. Alternative for North America: The connec-
tion may also be made to „Class 2 Circuits" or „Class 2 Power Units" according to CEC
(Canadian Electrical Code) or NEC (National Electrical Code).
Zero-point adjustment: Check the zero point shown on the display during commissioning.
If, for installation reasons, an offset is displayed, this can be reset in programming mode
using the 0SET parameter.

Wiring details

Ingress Protection per
IEC 60 529
26
© SICK AG • Germany • All rights reserved • Subject to change without notice
7. Starting, operation
Circular connector M12x1, 4-pin
2 switching outputs or
1 switching output +
1 analogue output
Q1 1 = 4
L+ = 1
M = 3
C/Q1 = 4
IP 67
The ingress protection classes specified only apply while the pressure transmitter is connected
with female connectors that provide the corresponding ingress protection.
Circular connector M12x1, 5-pin
2 switching outputs + 1 analogue output
Q2 = 2 /
L+ = 1
M = 3
Q
= 2
A
IP 67
PBS
Q1 = 4
Q2 = 2
Q
= 5
C/Q1 = 4
A
8013154/UP90/2015-03-30
Tabla de contenido
loading

Tabla de contenido