Electrical connection
Connect the instrument to earth via the pressure connection.
For power supply, use a circuit with energy limitation (EN/UL/IEC 61010-1, section 9.3)
with the following maximum values for the current: with L + = 35 V (DC): 4.0 A. Provide a
separate switch for the external power supply. Alternative for North America: The connec-
tion may also be made to „Class 2 Circuits" or „Class 2 Power Units" according to CEC
(Canadian Electrical Code) or NEC (National Electrical Code).
Zero-point adjustment: Check the zero point shown on the display during commissioning.
If, for installation reasons, an offset is displayed, this can be reset in programming mode
using the 0SET parameter.
Wiring details
Ingress Protection per
IEC 60 529
26