Operating Instructions
Mechanical connection
When mounting the instrument, ensure that the sealing faces of the instrument and the
measuring point are clean and undamaged.
Screw in or unscrew the instrument only via the flats using a suitable tool and the
prescribed torque. The appropriate torque depends on the dimension of the pressure
connection and on the sealing element used (shape/material). Do not use the case as
working surface for screwing in or unscrewing the instrument.
When screwing the transmitter in, ensure that the threads are not jammed.
The blue cap of the sensor housing can be twisted up to 330° on the stainless steel
housing until it can turn no more. Do not overtwist the cap.
With a flush-mounted process connection, remove the protective cap shortly before moun-
ting. Ensure that the membrane is not damaged during installation.