Emc / Emf Information - Philips InfraCare PR3110/00 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33
English
10
This symbol indicates that an object is capable of being recycled - not that the object has
been recycled or will be accepted in all recycling collection systems.
This symbol indicates that 'the device complies with European Medical Device Directive
93/42/EEC requirements'. 0344 refers to the notified body.
This symbol indicates the storage and transportation temperature limits to which the
+70ºC
infrared lamp can be safely exposed: -25°C to +70°C.
–25ºC
This symbol indicates: protected against access to hazardous parts with a finger or objects
of similar size and against vertically falling water drops.
This symbol warns you against optical radiation from the device which may cause eye
irritation.

EMC / EMF information

- This Philips device complies with all applicable standards and regulations regarding exposure to
electromagnetic fields.
- Use of the InfraCare device adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, the InfraCare device and the
other equipment should be observed to verify that they are operating normally.
- Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the InfraCare
device, including cables specified by the manufacturer. Otherwise, degradation of the performance
of this equipment could result.
Environment
The Philips InfraCare infrared lamp is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage power supply network that
supplies buildings used for domestic purposes.
Test levels
This device is compliant to 60601-1-2 Ed. 4 RF emissions Group 1 class B. RF emissions are very low
and are not likely to cause interference in nearby electronic equipment.
Immunity test levels
Electrical fast transient burst
Surge
Voltage dips short interruptions and voltage variations on
power supply input lines IEC61000-4-11
Power frequency magnetic field immunity test
IEC61000-4-8
Discharge level direct 8 kV, indirect 8
kV
2 kV
1 kV line to line
70% 25 cycles (50 Hz) 30 cycles (60
Hz)
30 A/m