5.4 Frequency measurement
To perform frequency measurement:
set the rotary switch at Hz%,
connect black test lead to COM terminal, and red test lead to
VΩ
Hz%CAPTemp terminal,
contact the tips of test probes to the points of measurement,
read the measurement result on the display.
See also sec. 6.3.2.
5.5 Measurement % of duty cycle (pulse filling indicator)
To perform the measurement:
set the rotary switch at Hz%,
press MODE button, until symbol % is shown on the display,
connect black test lead to COM terminal, and red test lead to
VΩ
Hz%CAPTemp terminal,
contact the tips of test probes to the points of measurement,
read the measurement result on the display.
5.6 Measurement of resistance
WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform measurement of resistance:
set the rotary switch at Ω
connect black test lead to COM terminal, and red test lead to
VΩ
Hz%CAPTemp terminal,
contact the tips of test probes to the points of measurement; the
best solution is to disconnect one side of the tested element, to
prevent the remaining part of the circuit interfere with the read-out
of the resistance value,
read the measurement result on the display.
,
CMP-3000 – USER MANUAL
55