3.3

Voltage measurement 1P

During the single-pole determination of
phases to mark the external conductors,
in some conditions the operation of the
tester may deteriorate (e.g. when using
isolated personal protection equipment or
isolated stand).
Single-pole phase tests should not be
considered as sufficient to determine
whether the circuit is energized. 2P (2-
pole) tests should be performed in this
case.
When operating in this mode, the Auto
OFF function of the tester is inactive.
Performing the measurement is possible
in protective gloves without the use of a
touch electrode.
 Firmly grasp L2 probe in the
area between the conductor
and the protective barrier.
 Connect probe L2 of the test-
er to the tested object.
The presence of AC voltage is
indicated by active indicator
lights.
 After disconnecting the volt-
age, the tester switches off
after a few seconds.
P-4 ● P-5 ● P-6 – USER MANUAL
AC voltage indicator light
27