Display Backlight - Tintometer Lovibond MD 100 Instrucciones

Ocultar thumbs Ver también para Lovibond MD 100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 94
Functional description
GB
Zero
Test
Zero
Test
Zero
Test
Sample
Zero
Test
!
34
Press[ZERO/TEST]button
Removethe vial from the sample chamber
OTZ (One Time Zero)
The One Time Zero is available for all photometer variants whereby
the zero adjustment is carried out in a 24-mm round vial with sample
water. It canbe usedif different tests are carried out with identical test
conditions and the same water sample. When changing methods, it
is not necessaryto carry out another zero adjustment; the next test
can begin immediately. The zero adjustment is saveduntil the device
is switched off.
The zero setting can be repeated each time if necessary.
One Time Zero(OTZ),1x performing zero, then parameter measurement
Repeating the zero,
Pressthe key for 2 seconds.
Performing test procedure
Add reagent (f.e. a reagent tablet) (seemethod description)
Close vial
Placesample vial in the sample chamber.
Payattention to the positioning.
Press[ZERO/TEST]button

Display backlight

press,the backlight is switched off automatically during the meas-
urement
MD1X0/200_1b 09/2020
Tabla de contenido
loading

Este manual también es adecuado para:

Lovibond md 110Lovibond md 200

Tabla de contenido