SonoSite DICOM Guia Del Usuario página 7

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 34
2 On the Connectivity setup page, select 
DICOM Setup.
3 Select Config.
4 From the Configure list, select a configuration 
page.
5 Select New.
6 Fill in the fields. Select Next if present, and fill 
in the fields on page 2.
See one of the following:
• "Location configuration (page 1)" on 
page 3
• "Archiver configuration (page 1)" on 
page 5
• "Printer configuration (page 1)" on page 5
• "Storage Commit configuration" on 
page 6
• "Worklist configuration (page 1)" on 
page 6
• "MPPS configuration (page 1)" on page 6
To enter special characters:
• (M‐Turbo) Select Symbols.
• (S Series) Click Symbols on the on‐screen 
keyboard.
For more information, see the ultrasound 
system user guide.
To enter an underscore, use the 
(Underscores are disallowed in the Host 
Name field.)
To undo the last change, select Cancel.
7 (Printer or Wireless Location Only) Select 
Next, and fill in the fields on Page 3.
8 (Wireless Location Only) If you selected 
RADIUS for Security Policy or for WPA
Authentication, select Next and fill in the 
fields on page 4.
9 Select Save and then select Done.
10 Complete any additional configuration tasks, 
and then select Done.
A dialog box prompts you to restart the 
system.
To delete a location or device
1 On the proper configuration page, select the 
name from the list of locations or devices.
2 Select Delete.
3 Select Yes to confirm deletion.
4 Complete any additional configuration tasks, 
and then select Done.
A dialog box prompts you to restart the 
system.
Configuration setup fields
Location configuration (page 1)
Host Name Unique network name for the 
ultrasound system. Default is SonoSite.
Alias Name used to identify the network location 
of the ultrasound system.
AE Title DICOM Application Entity Title.
Wireless Select if configuring a wireless location.
Port Device port number. IP port 104 is typically 
assigned for DICOM.
SPACEBAR
DHCP If desired, select to enable Dynamic Host 
Configuration Protocol. The IP Address, Subnet 
Mask, Default Gateway, and Alternate Gateway 
fields become unavailable.
IP Address Unique identifier of the ultrasound 
system location. Cannot be between 127.0.0.0 
and 127.0.0.8.
Subnet Mask Identifies a network subdivision. 
The default value is 255.255.0.0.
3
Tabla de contenido
loading

Tabla de contenido