Error displayed
Cause
Password incorrect
Entry invalid!
Data analysis not
No valid data for these
possible, no
parameters!
measurement data
View Data not possible in
No valid data found!
data log
No help function
present.
Data analysis settings
cannot be configured
No measurement data
without measurement
present!
data.
Data analysis limits not
Control range not
reached
reached!
Data analysis limits
Control range
exceeded.
exceeded!
Calculated concentration
Concentration too high!
is higher than 999999
The measured
absorbance is above the
Over measuring range
calibration range of the
test
The measured
absorbance is below the
Under measuring range
calibration range of the
test
Possible interference
Interference Check
by:
English 28
Resolution
Error displayed
Forgotten your
Possible interference
password?
from:
Contact technical
support.
Next service is due!
Change the selection.
Negative result!
Change the selection.
Unstable lighting
conditions!
System check
incorrect!
Change the selection.
This is a warning notice.
Temperature too high.
The control limit set was
Measurement not
not reached.
possible!
This is a warning notice.
The control limit was
exceeded.
Update errors
Dilute sample and
An error occurred
measure again
when uploading the
instrument data.
Dilute sample and
measure again
An error occurred when
reading from the USB
memory stick.
If possible, select a test
with a lower reading
An error occurred when
range or use a cuvette
writing to the USB
with a longer path length
memory stick.
The analysis is possibly
erroneous due to
Please check on the
interferences.
current update file.
Please contact
Customer Service.
Cause
Resolution
The analysis is possibly
Interference Check
erroneous due to
interferences.
Contact technical support
for an inspection of the
instrument.
The calculated result is
Check concentration of
negative
sample
Avoid direct sunlight at
the measuring location.
Switch the instrument off
and then back on again.
Measurement of air
If the system check is not
values failed
successful, contact
technical support.
Switch of the instrument
and allow it to cool for a
few minutes. If
necessary, move it to a
cooler place.
Start the procedure again
or contact technical
support.
Start the procedure again
or contact technical
support.
Start the procedure again
or contact technical
support.
Check the USB memory
Error during update.
stick.
Error during update.