TOTALIZATION
Totalization measures the cumulative weight of a
sequence of items. (See Figure 62) The cumulative
total may exceed the capacity of the balance. The
maximum number of samples (n) is 99.
Totalization
Place sample on the pan. Press Accumulate to add to the total.
70.19
>0<
Gross:
70.19 g
Samples:
Tare:
0.00 g
Total:
Result
Clear Total
Accumulate
Figure 62
1.
In the upper left portion of the home screen, select
Totalization.
2. Place item on the pan to begin. The sample weight
is shown on the main display line.
3. Press ACCUMULATE to add the weight (when stable)
of the item to the total.
4. Remove the item from the weighing pan, then add
the next item and continue as above.
5. Press RESULT to view the results from the totalization.
(See Figure 63)
Figure 63
Totalization
Samples
Average
o (stdev)
Minimum
Maximum
Save to USB
6. When finished, press the Clear Total button to reset
the accumulated weight to zero.
7.
Press SAVE TO USB to save the result to a USB flash
drive or EXIT to return to the totalization home screen.
NOTE: Changing units converts the
accumulation results to the selected unit.
PAGE 16 OF 54
OPERATION CONTINUED
The
Max 4200 g
d=0.01 g
TOTALIZATION
home screen
Main Display Line
g
0
Reference Fields
0.00 g
Function Buttons
Menu
Item
Result
3
Total
506.95 g
168.98 g
38.90 g
117.00 g
210.57 g
Range
93.57 g
Exit
FORMULATION
Use this application for compounding and recipe
making. (See Figure 64) The maximum number of
components is 50.
Formulation
Press Start to begin Formulation.
0.00
>0<
Gross:
0.00 g
Filler:
Tare:
0.00 g
Auto Tare:
Item
Start
Settings
Figure 64
1.
In the upper left portion of the home screen, select
Formulation. (See Figure 65)
2. Press START to begin the compounding process.
3. Place the first ingredient on the pan (or in a
tared container) and press ACCEPT to store the
component.
4. Continue adding components and pressing ACCEPT
to store the weight of the individual components
until the formula is complete. The Total line shows the
total weight of all the components.
5. Press STOP to finish the formulation. The formulation
results are displayed:
NOTE: If Filler is set to On (see Item Settings on
page 17), the balance will ask to add a filler
material to complete the formulation. Add the
filler material and press ACCEPT to complete
the formulation and display the results.
Figure 65
Formulation
Comp.
Comp. Total
Save to USB
6. Press SAVE TO USB to save the results to a USB flash
drive or EXIT to return to the Formulation Home screen.
NOTE: The formulation results will be cleared
when a new formulation starts.
The
Max 4200 g
d=0.01 g
FORMULATION
home screen
Main Display Line
g
Off
Reference Fields
On
Function Buttons
Menu
Weight
1
241.76 g
2
272.95 g
514.71 g
Exit
0521 IH-5276