7. Sterilization
Sterilize the motor only.
For the sterilization method, we recommend the autoclave sterilization
method.
Sterilization is required first time you use and after each patient as noted
below.
Autoclaving
) Turn off the Power.
2 ) Detach the motor from the Motor Cord. ( Refer to 4- Connecting /
disconnecting the Motor and the Motor Cord)
3 ) Clean the surface of the motor with brush etc. ( Do not use the
metal brush ) , and wipe it with the cotton moistened with
disinfecting alcohol.
4 ) Screw the Motor Cap to the Motor. Put the Protection Plug to the motor insert. ( Fig. 0 )
5 ) Insert into an autoclave pouch. Seal the pouch.
6 ) Autoclavable up to max. 35°C.
ex. ) Autoclave for 20 min. at 2°C, or 5 min. at 32°C.
7 ) Keep the Motor in the autoclave pouch to keep it clean until you use it.
Sterilization at 2°C for more than 5 minutes is recommended by EN3060 or EN ISO7665-.
CAUTION
· Do not Autoclave (or any other high temperature Sterilization) Control Box, AC Adapter, Motor Cord.
· Do not lubricate the motor.
· Do not wipe nor immerse the system in acidic water or acidic solutions.
· Do not sterilize with dirt on the surface. It might cause rust.
· Do not use Protection Plug without the mounting O-Ring. It may cause a malfunction. If O-Ring had damaged,
replace it immediately.
O-Ring (Protection Plug ): Order code 03245702
· You can hang Protection Plug on the anywhere to prevent loosing it.
· Do not hung the Protection Plug with motor. It may drop the motor and cause damage.
8. Maintenance
If the O-Ring became depleted, such as difficult to connect the
handpiece or air or water might leak, replace the O-Ring.
() Changing the Oring
Remove the O-Ring at the Motor Insert with a pointed tool, and
mount the new O-Rings into the groove.
CAUTION
There are 4 O-Rings on the Motor Insert. Blue one is thinner than other 3 pieces.
Make sure to order it.
Protection Plug
8
Motor Cap
Tighten
Fig.0
Fig.
O-Ring ( Blue )
O-Ring ( Black )