NSK Ti-Max NLX nano Manual De Operación página 13

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Guidance and manufacturer's declaration - electromagnetic immunity
The NLX nano is intended for use in the electromagnetic environment specified below. The customer or the user of the NLX nano and manual should assure that it is
used in such an environment.
Immunity test
IEC6060 test level
Conducted RF
3Vrms
IEC6000-4-6
50 kHz to 80MHz
Radiated RF
3V/m
IEC6000-4-3
80MHz to 2.5 GHz
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the NLX nano is used exceeds the applicable RF compliance
level above, the NLX nano should be observed to verity normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the NLX nano.
b Over the frequency range 50kHz to 80MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the NLX nano
The NLX nano is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the NLX nano can
help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.0
0.

0
00
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Compliance level
3Vrms
3V/m
Separation distance according to frequency of transmitter
50kHz to 80MHz
0.2
0.38
.2
3.8
2
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any
part of the NLX nano, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = .2 P
d = .2 P 80MHz to 800MHz
d = 2.3 P 800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in watts
(W)according to the transmitter manufacturer and d is the recommended
separationdistance in meters (m).
Field strengths from fixed RF transmitters as determined by an electromagnetic
a
site survey
, should be less than the compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with the following
symbol:
80MHz to 800MHz
0.2
0.38
.2
3.8
2
2
b
,
800MHz to 2.5GHz
0.23
0.73
2.3
7.3
23

Capítulos

Tabla de contenido
loading

Tabla de contenido