4. Handling the Motor
4-1 Connecting / disconnecting the motor and the motor cord
Align and insert firmly the Motor Pin into the pin holes of the Motor
Cord Connector, and fasten the Motor Cord Nut securely. (Fig. 5)
To remove the Motor Cord from the Motor, unscrew and detach the
motor cord nut, and gently pull out the motor cord connector.
4-2 Connectioning / disconnecting the motor and the handpiece
To insert the E-type handpiece in the Motor Insert, align handpiece
and motor, then turn those until its click. (Positioning pins are
aligned)
To remove the handpiece, simply pull out the handpiece from the
motor.
Handpiece (not included in NLX nano set.)
CAUTION
· Care should be taken when using any handpiece attachment which requires lubrication (E Type Micromotor).
Following lubrication stand the handpiece attachment vertically until the excess solution drains. Wipe the
handpiece then attach it to the Micromotor.
· Do not connect or disconnect the handpiece until the motor has completely stopped.
5. Operation Section
5- Insert the Power Cord Plug in AC outlet and turn the Power Switch on. When you turn on the
Power Switch, lamps and Indicator will come on.
5-2 Before using this system, make sure to do the Foot Air Calibration function.
( See 6- Foot Air Calibration )
5-3 Push the Gear Ratio Select Key to indicate the speed of the
attachment you will use. (Fig. 7)
During Speed Setting, Indicator display "Speed".
For detail display, check table below.
Speed Display Table
Handpi e ce
Indicator
Gear Rati o
16:1
1 - 25
1:1
2 - 40
1:5
1 - 20
Speed ( min
(r pm ))
-1
100 - 2,500
2,000 - 40,000
10,000 - 200,000
5
Motor Cord Connector
Motor
Motor Cord Nut
Handpiece (Option)
Positioning Pin
Motor Insert
Power Switch
Symbol Mark
Function
Gear Ratio Lamp
Motor Cord
Fig.5
Fig.6
OFF
ON
PUSH
Fig.7