MANUAL DE INSTRUCCIONES CODIGO 80098 REV E
ID number
Temperature
offset
Temperature
coefficient
Serial outlet
type
Logging
interval
Fig 1. Display:
J
.P. SELECTA s.a.
May/2017
7.5.1 Setting «Offset»
The «Offset» allows adding a quantity (positive or negative) to the measured tem-
perature value, moving the reading.
An «offset» between -9.9 and 9.9 can be set.
a. The «RUN» led must be off (press «START/STOP»).
b. Press
c. Release
d. Press
e. Press
f.
Press
7.5.2 Setting the «Temperature coefficient»
The temperature coefficient adds a factor that multiples the controller original reading.
A coefficient between 800 and 1000 can be set (The value 1000 is shown in the
display as 000)
The coefficient default value is 1000.
This coefficient applies according to the formula:
Tª_display = Tª_probe x (coefficient / 1000).
a. The «RUN» led must be off (Press «START/STOP»).
b. Press
c. Release
d. Press
e. Press
f.
Press
7.6 Setting the safety thermostat
DRY-BIG ovens are equipped with a safety thermostat (See Fig. 4). This device
interrupts the oven's heater if it exceeds the set value.
It is intended as an additional safety device, in case the temperature electronic con-
troller stops operating properly. In this case, its action prevents that the temperature
could rise to excessive values.
Another use of the thermostat is to protect valuable samples, by setting it between
5 and 10 ºC above the operating temperature.
The thermostat comes factory set to maximum: 100% (at full clockwise).
To set the safety thermostat shutdown temperature, follow the next instructions:
a. Set the thermostat at full clockwise (100%).
b. Be sure that the safety thermostat is reset by pressing «RESET». The red lamp
must be off (See Fig. 4).
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso)
and without releasing, press
and
.
once.
or
until the display shows the desired «Offset».
four times.
and without releasing, press
and
.
twice.
or
until the display shows the desired «coefficient».
three times.
Pag.: 13
for 5 seconds.
for 5 seconds.
Tel 34 93 770 08 77 Fax 34 93 770 23 62