Technical Data; Components; Preassembled Systems; Conex - Grundfos Conex DIA-1-A PA Instrucciones De Instalación Y Funcionamient

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 34

8. Technical data

8.1 Components

8.1.1 Measuring cells
Measuring cell for peracetic acid
Peracetic acid
(covered with diaphragm), including
measuring cell
flow armature and cables
Measuring cell for hydrogen peroxide
Hydrogen peroxide
(covered with diaphragm), including
measuring cell
flow armature and cables

8.1.2 Preassembled systems

®

Conex

DIA-1-A, PA/HP-PA (Conex
Peracetic acid
measuring cell
Instrument amplifier
For wall-mounting, mounted on a
and controller
PVC plate
®
Conex
DIA-1
For panel-mounting
®
(Conex
DIA-2Q)
®
Conex
DIA-1-A, PA/HP-HP (Conex
Hydrogen peroxide
measuring cell
Instrument amplifier
For wall-mounting, mounted on a
and controller
PVC plate
®
Conex
DIA-1
For panel-mounting
®
(Conex
DIA-2Q)

8.2 General data

PVC, polycarbonate, stainless steel and
silicone rubber, resistent against
Enclosure
tensides and comparable water
additives
Standard ranges 0-100 / 0-500 / 0-1000 /
Measuring ranges
0-2000 mg/l or freely adjustable within
0-2000 mg/l
Permissible pH range pH 1 to 11
Recommended
Minimum 30 l/h
sample-water flow
Measuring signal is
Temperature drift
temperature-compensated
Peracetic acid measuring cell
Accuracy
2 % of measured value ± 5 ppm
Repeatability
2 % of measured value ± 3 ppm
No disturbances by tensides
Disturbances
Negligible disturbance by hydrogen
peroxide, selectivity factor = 0.005
Resolution
1 mg/l
Response time, T
Approximately 3 minutes
90
Permissible working
0 °C to 50 °C
temperature
8
®
DIA-2Q-A, PA/HP-PA)
®
DIA-2Q-A, PA/HP-HP)
Hydrogen peroxide measuring cell
Accuracy
2 % of measured value
Repeatability
2 % of measured value
No disturbance by tensides
Disturbances
Strong disturbance by peracetic acid
Resolution
1 mg/l
Response time, T
Approximately 5 minutes
90
Permissible working
0 °C to 45 °C
temperature
®
Voltage for Conex
DIA-1 (Conex
Type designation
Description
DIA-1 (-2Q),
230/240 V (50/60 Hz)
1-P/R/D/HP/PA/F, W-G
DIA-1 (-2Q),
115/120 V (50/60 Hz)
1-P/R/D/HP/PA/F, W-H
DIA-1 (-2Q),
24 V (DC)
1-P/R/D/HP/PA/F, W-I
®
DIA-2Q)
Tabla de contenido
loading

Este manual también es adecuado para:

Conex dia-1-a hpConex dia-2q-a paConex dia-2q-a hp

Tabla de contenido