Fault
Visual alarms flashing, but
no acoustic alarm and no
audio response.
Acoustic alarm, but no
visual alarm flashing.
Acoustic alarm sounds
and all visual alarms
flashing.
Alarm: Device
malfunction.
No audio response.
MEDUMAT Easy CPR is
functioning, but without
any displays.
V
too low.
t
Unusually high oxygen
consumption.
MEDUMAT Easy CPR will
not switch off.
Pressure gauge not
reading "0".
The MEDUtrigger does
not trigger a respiratory
stroke (CPR mode only)
The MEDUtrigger does
not work
EN
60
Troubleshooting
Cause of fault
Short-term electronic problem or
defect in electronic system.
Device defective.
Audio response deactivated.
Pressure gauge hose on
MEDUMAT Easy CPR or on
patient valve slipped off.
Kink in pressure gauge hose.
Ventilation parameters incorrectly
set.
MEDUMAT Easy CPR defective.
Leak in oxygen supply.
Operating error.
MEDUMAT Easy CPR defective.
Respiratory stroke was triggered
during the inspiration phase
MEDUtrigger not correctly
connected
Remedy
Switch off, then on again.
If error recurs, arrange for repairs.
Activate audio response (5.11,
page 35).
Check pressure gauge hose.
Check ventilation parameters.
Have repaired.
Find and eliminate leak (7.3, page 49).
Keep key pressed for at least 3 seconds.
Arrange for repair.
Wait until the expiration phase is over
to trigger the respiratory phase
(corresponds to the length of the
inspiration phase)
Check whether the MEDUtrigger plug
is correctly positioned and correct it