Lovibond Tintometer MD 100 Manual De Instrucciones página 52

Ocultar thumbs Ver también para Tintometer MD 100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 183
10 ml
24 mm
Fill 24 mm vial with 10 ml
prepared sample .
Close vial(s).
Test
Press the TEST (XD:
START)button.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/l total Iron or when using a filtrated sample, in mg/l totale soluble Iron
appears on the display.
Implementation of the provision Iron LR (A) with liquid reagent
Select the method on the device
For determination of total dissolved iron the sample must be filtered prior to the test
(pore size 0,45 μm). Otherwise, iron particles and suspended iron are measured.
EN   Method Reference Book   1.0
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Invert several times to mix
the contents.
5 min
12
11
1
10
2
9
3
8
4
7
5
6
Wait for 5 minute(s)
reaction time.
Iron LR L (A) / M225
Add 10 drops
KS61 (Ferrozine/
Thioglycolat).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
10
EN
53
Tabla de contenido
loading

Este manual también es adecuado para:

Tintometer md 110Tintometer md 200

Tabla de contenido