Contra angle lubrication

Lubricate the contra angle head only by using the X-SMART
Insert the spray nozzle into the rear part of the contra angle head and lubricate the
head for 2-3 seconds.
CAUTION:
Do not lubricate the motor section of the motor handpiece.
To clean the motor handpiece, do not use any solvent such as benzene and/or a
thinner.
Before mounting the lubricated contra angle to the motor handpiece, wipe off extra
oil. Stand it on its end or lean it in the proper position for gravity draining. Mount it
after excess oil has been drained.
Hold the contra angle securely to prevent it from detachment caused by the
pressure of the spray.
Supply lubricant until it comes out of the handpiece head (for approx. 2 seconds).
Do not use a spray can upside down as only spray gas will be dispensed without
oil.
NVG / F190294X / 06 / 2007 –updated 08/2009
Operating mode
- Immerse the accessories in a
disinfectant solution (bactericidal,
virucidal, fungicidal,
tuberculocidal and aldehyde free)
according to the manufacturer's
recommendations
- See point # 2
- After rinsing, the accessories
must be dried.
- Inspect devices and sort out
those with defects.
- Dirty instruments must be
cleaned and disinfected again.
- Pack the devices in "Sterilization
pouches".
- Steam sterilization at:
134°C (274°F) for 18 min
- Keep devices in sterilization
packaging in a dry and clean
environment
Warning
- Follow instructions and observe concentrations
and suggested times given by the manufacturer
- Check the validity period of the pouch given by
the manufacturer to determine the shelf life.
- Use packaging which is resistant to a
temperature up to 141°C (286°F) and in
accordance with EN ISO 11607.
- The Lip clip, Hook, Contra angle and its
connection parts with contact fork must be
sterilized according to the packaging labelling.
- Use fractionated vacuum or (less preferred)
gravity autoclaves (according to EN 13060, EN
285).
- Use validated sterilization procedure according
to ISO 17665-1
- Use the maintenance procedure of the
autoclave device given by the manufacturer.
- Use only the listed sterilization procedures.
- Sterility cannot be guaranteed if packaging is
open, damaged or wet (check the packaging
before using the instruments).