VATECH EzRay Air Wall VEX-S300W Manual Del Usario página 44

Ocultar thumbs Ver también para EzRay Air Wall VEX-S300W:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

4. Troubleshooting
Troubleshooting
Problem
Equipment is not
turned on.
Control Panel is not
turned on.
No X-ray emission
X-ray emission
works, but
exposure is too light
completely white.
X-ray emission
works, but
exposure is too
dark.
36
Power button is not turned
on properly.
Defective main board
Internal cable disconnected
Generator is cooling.
Remote Exposure Switch is
pressed, but X-ray
exposure does not work.
Internal cable disconnected
Defective generator
Tube lifecycle termination
Equipment has been
positioned incorrectly.
Exposure time is too short.
or
Receptor is facing the
wrong way.
Exposure time is too long.
Cause
Turn the equipment power
switch off and turn it back on.
Contact your Service
Representative.
Contact your Service
Representative.
Wait for the cooling time.
(Duty Cycle: 1:60)
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Adjust the position of the
equipment.
Increase the exposure time.
Reposition the receptor.
Decrease the exposure time.
EzRay Air Wall (VEX-S300W) User Manual
Solution
Tabla de contenido

Solución de problemas

loading

Tabla de contenido