Electric Dermatome Models B and PI
Cleaning and Decontamination (continued)
into the oscillating pin recess of the cutting head. Transport the tray containing the soiled
dermatome in an impervious plastic bag or container to the decontamination environment.
Keep the outside of the container clean.
2. Enzymatic Soak: Prepare a neutral pH enzymatic detergent, ASP® ENZOL®, as per the
manufacturer's recommendation at 1 oz. per gallon (~8mL per Liter) using lukewarm
tap water. Remove the four #4-40 x .38L stainless steel flat head screws fastening the
dermatome's cutting head to the motor housing. Immerse only the dermatome's cutting
head, width clips, guard, screw fasteners, screwdriver, and pin position guide in the
enzymatic solution. Soak for a minimum period of 30 minutes. Do not immerse the
dermatome's motor housing. Immersion of the motor housing can cause permanent
damage to its electrical components.
CAUTION: Removing the dermatome's cutting head from the motor housing will result
in exposing the oscillating pin and motor drive shaft. The contour of the oscillating pin is
factory set to provide proper dermatome performance. Care must be taken in handling the
motor housing in this state. Dropping the motor housing or hitting the oscillating pin can
cause damage that may adversely affect dermatome performance.
3. Rinse: After the enzymatic detergent soak period, remove dermatome's cutting head, width
clips, guard plate, width clip screws, #4-40 x .38L flat head screws, dermatome screwdriver,
and pin position guide from the detergent solution and thoroughly rinse the items with
lukewarm tap water.
4. Clean Instruments: Prepare a neutral pH enzymatic detergent, ASP ENZOL, as per the
manufacturer's recommendation at 1 oz. per gallon (~8mL per Liter) using lukewarm
tap water. Using a small clean soft-bristled brush, remove soil from all surfaces of the
dermatome's cutting head, width clips, guard plate, width clip screws, #4-40 x .38L flat head
screws, dermatome screwdriver, and pin position guide while they are submerged in the
solution. Take particular care in removing soils from the dermatome's cutting head features
containing crevices and sharp inside corners. Clean the dermatome's motor housing by
wiping with a soft clean cloth moistened with the detergent solution. Pay special attention
to removing soil from the area of the oscillating pin and drive shaft. Clean the handpiece
cable by wiping with a soft clean cloth moistened with the detergent solution.
5. Rinse: Thoroughly rinse all dermatome components under gently running lukewarm tap
water and wipe with a dry, soft, clean cloth. Do not immerse the dermatome components.
Dry all components with a soft, dry towel.
6. Final Rinse: Rinse all dermatome components with softened or deionized water. Softened
or deionized water should be used for the final rinse to better remove detergents and other
residues from the rinse water. Excessively hard water can spot or stain instruments. Excess
chlorine in the rinse water can result in pitting. Dry all components with a soft, dry towel.
19
®