Chlorine PP / M110
Chemical Method
DPD
Appendix
Interferences
EN
Persistant Interferences
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
Concentrations above 2 mg/l Chlorine, in the event of using Powder Packs, can lead
to results within the measuring range of up to 0 mg/l. In this case, the sample must be
diluted with chlorine-free water. 10 ml of the diluted sample should be mixed with the
reagent and the measurement taken again (plausibility test).
Interference
CrO
2-
4
MnO
2
Method Validation
Limit of Detection
Limit of Quantification
End of Measuring Range
Sensitivity
Confidence Intervall
Standard Deviation
Variation Coefficient
Conformity
EN ISO 7393-2
determination of free, combined and total
a)
34
from / [mg/l]
0,01
0,01
0.01 mg/l
0.03 mg/l
2 mg/l
1.68 mg/l / Abs
0.033 mg/l
0.014 mg/l
1.34 %
EN   Method Reference Book   1.0