2 Shut down the density meter.
3 Press on the grip recesses (4) of the cover and lift the
lower part of the cover out of the housing.
4 Pull the top of the cover out of the housing.
5 Lift the piston (1) out of the guide system.
6 Pull the cylinder (2) up and out of its guide system (3).
Install the sample pump
1 Insert the cylinder (3) of the sample pump in its guide
system (4).
2 Insert the piston (1) in its guide system (2).
3 To check if the sample pump works, move the navigation
joystick downwards and let go.
ð The sample pump starts and stops when the piston is
in the lowermost position.
4 Reinstall the cover of the sample pump.
6.6 Prepare the density meter for storage
The density meter can be stored at temperatures of -20...+50 °C (-4...+122 °F). But the lifespan of the battery
is reduced when it is stored at high temperatures.
1 Clean the measuring cell.
2 Shut down the density meter.
3 Clean the housing.
4 Store the density meter in a dry and clean place.
See also
2
Clean the density meter } Page 15
2
Technical data } Page 20
6.7 Dispose of the density meter
In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to
countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment. If you have any questions, please contact the
responsible authority or the distributor from which you purchased this device. Should this
device be passed on to other parties, the content of this regulation must also be related.