Braun ThermoScan IRT 6515 Manual Del Usuario página 8

Ocultar thumbs Ver también para ThermoScan IRT 6515:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Fit the probe snuggly into the ear canal, then push
4
and release the Start button
ExacTemp light will pulse while temperature measurement is in progress. The
5
light will remain solid for 3 seconds to indicate that a successful temperature
reading has been achieved.
NOTE: If the probe has been properly inserted into the ear canal during the
measurement, a long beep will sound to signal the completed measurement.
If the probe has NOT been constantly placed in a stable position in the ear
canal, a sequence of short beeps will sound, the ExacTemp light will go out
and the display will show an error message (POS = position error).
See «Errors and troubleshooting» section for more information.
The confirmation beep indicates that an accurate temperature measurement has been taken.
6
The result is shown on the display. Temperature reading displayed is an equivalent of oral
temperature.
To repeat
Go to step 3.
For the next measurement, press Eject button
7
remove and discard used lens filter, and put on a new,
clean lens filter.
The Braun ThermoScan ear thermometer turns off
automatically after 60 seconds of inactivity.
The thermometer can also be turned off by
pressing the Power button
The display will briefly flash OFF and it will go blank.
6
.
to
.
Tabla de contenido
loading

Tabla de contenido