SPARE PARTS/ACCESSORIES

Applied part
Item description
®
INQUA
nebuliser set for INQUA

TECHNICAL DATA

General

Recommended nebuliser gases:
Minimum flow:
Corresponding operating pressure:
Maximum flow:
Corresponding operating pressure:
Minimum fill volume:
Maximum fill volume:
Aerosol characteristics in accordance with
EN 13544-1 appendix CC
The output of the nebuliser depends on the medication being used.
Consequently, the following data may not apply for medications that
exist as suspensions or are very viscous. If in doubt, ask the manu-
facturer of the medication concerned for advice.
Determination of particle size
®
Fill the INQUA
attach the INQUA
tion Impactor™) and run for 3 minutes.
nebuliser with 3 ml of a 2.5% test solution and then
®
nebuliser to an impactor (NGI™ = Next Genera-
®
Inhalator 1000
Air
(3.0 l/min)
50 kPa (0.5 bar)
(6.0 l/min)
200 kPa (2.0 bar)
2 ml
8 ml