Starting Up; Equipment Usage - Selecta HPC PL-3 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES
Door open
indicator
Fig 1. Starting up screen
Fig 2. Program selection screen
Fig 3. Program running screen
J
.P. SELECTA s.a.u.
CODIGO 80394 REV 0

6.2 Starting up

Before starting up, please be sure that:
- You have read the instructions referred to in paragraph 5 «Installation».
- Activate the main switch (5) located at the rear part of the equipment.
- All actions are carried out through the control screen (2).
- When starting the equipment up, the work screen will appear. This screen indicates
that the door is open and you will not be able to start the cycle until it is closed. When
you open the door, the fan will stop heating and stirring.
An icon also appears to be able to download the recorded data from the previous
process.

6.3 Equipment usage

Once the main screen starts up, it turns on and the button for START appears on screen.
The equipment has two operating modes:
- The standard functionality for most "1-3 BAGS" applications
- The extra operation to place up to 6 bags "4-6 BAGS" (This process is only recom-
mended for exceptional cases of high load and what it does is two heating sequences.
At the end of the first phase it will be necessary to remove the bags from the top tray
and continue the process with the bottom tray).
- Once the process has started, it cannot be stopped to prevent it from accidentally
stopping. It can only be stopped if the equipment is turned off with the main switch (5).
-While the bags show freezing signs, the ice symbol appears and the temperature
reading will not be displayed. The reading will only be shown after 25ºC.
- At the end of the process it will indicate END OF PROCESS.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - http://www.grupo-selecta.com
04/2020
Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
Pag.: 11
Tabla de contenido
loading

Tabla de contenido