Step 8:
If the Genius™ 3 Thermometer is within specified accuracy limits, a Results screen will be displayed showing the thermometer serial number along with an
indication that the thermometer passed calibration check. If the Genius™ 3 Thermometer is found to be outside the specified accuracy limits , the display
will indicate to the user that the Genius™ Checker/Calibrator is entering into a calibration sequence that requires the user to repeat Steps 4 through 7
three more times in order to gather recalibration data. Following the last pass through Steps 4 through 7, the Genius™ Checker/Calibrator will attempt to
recalibrate the Genius™ 3 Thermometer. If the recalibration is successful, a Results screen will be displayed showing the thermometer serial number along
with an indication that the thermometer passed calibration. If the recalibration is unsuccessful or the thermometer is unable to be calibrated, a Results
screen will be displayed showing the thermometer serial number along with an indication that the thermometer failed calibration.
Step 9:
After the pass/fail Results screen has been displayed, press the encoder knob to display a second Results screen containing the target temperatures, the
Genius™ 3 Thermometer readings, and the target versus thermometer variances.
Step 10:
The next screen will display "Write report to USB drive?" and display Yes and No below. If a report is desired, insert a flash drive into the front panel USB
port. Note: allow about 30 seconds after insertion for the USB flash drive to initialize. Use the encoder knob to scroll to either Yes or No and press the
encoder knob to make a selection. If No is selected, the screen returns to the Main Menu. If Yes is selected, a report is written to the USB flash drive, using
the thermometer serial number for filename and .TXT as the file extension. The file contains the test results as well as other information such as serial
number, date, time, user name, and organization. This file is a simple ASCII text file and is only available in English. This file can be read by Microsoft
Windows® operating system and Microsoft Windows® notepad and sent to a printer. After the file is written, the display will show "Push knob to continue."
Press the encoder knob to return to the Main Menu.
If the thermometer fails recalibration, reattempt calibration by repeating the Check Calibration steps. If the thermometer persistently fails recalibration,
the thermometer should be returned for service.
Step 11:
After completing a calibration check/recalibration, remove the probe connector from the battery compartment by pulling straight back on the connector
and rotating up and out of the thermometer as shown in Figure 4. Then reinstall the 3 AAA batteries, observing the proper polarity orientation. Finally,
reinstall the battery compartment cover and tighten the screw or screws.
Step 12:
Remove USB flash drive and power connector from the Genius™ Checker/Calibrator front panel before closing the lid. The lid should be closed when the
Genius™ Checker/Calibrator is not in use.
Genius™ 2
Genius™ 3
Figure 4: Probe Connector removal