Taking measurements using WatchBP Office

«AUSCULTATION» Mode
1) Turn on the power – Turn on the device by
switching the I/O switch at the back of the
device into the ON position and make sure
Bluetooth switch is in off position.
DC 6V 600mA
2) Apply the cuff* – Properly fit one cuff to the
preferred measurement arm.
Additional visual instruction can be found on the cuff
*
16
3) Switch to «AUSCULTATION» Mode – Slide the
Mode Switch to «AUSCULTATION» Mode.
AUSCULT 1st FOLLOW UP
4) Adjust inflation pressure if necessary –
Inflation pressure can be set manually from 140
to 280 mmHg by pressing and holding the Start/
Stop Button for 7 seconds.
Press the M Button to adjust the pressure and
press Start/Stop to save the setting.
AUSCULT 1st FO