Maintenance to be carried out by the surgery staff
(fi g. 2-3-7/A-7/B-8-10 Page 51-52-55-56-57-58)
Biological danger, risk of infections from epidemic diseases
General danger sign
High temperature
Before servicing, follow the instructions given at : " Signals and warnings" and "Notice
and biological danger". After each surgical operation or after lengthy treatments,
rinse the system thoroughly by aspirating large quantities of clear water. At the end
of each working day, clean and sanitize the system by aspirating a solution
of warm (50 °C) water and Puli-Jet, prepared according to the instructions.
Replace dirty fi lters (fi g. 7/A-7/B-8-10 det. A). Aspirate the solution by means of
"Pulse Cleaner", which will create the necessary turbulence to optimise cleansing
and to drain off the debris.
In the morning, before starting to work, make sure that antifoaming sanitizing tablets
are placed in the tip support fi lter. Using products other than those recommended by the
manufacturer may affect the functioning of the system and cause failure.
Clean or replace the secretion fi lter (indicated at pages 51 and 52 with the arrow
"secretion inlet" and at pages 57 and 58 with detail A). Any possible amalgam present
in the fi lters must be collected in accordance with ISO norm 11143 and delivered to
authorised disposal companies, as this is a toxic, harmful and contaminated waste.
Maintenance and periodical inspections to be carried out by
authorised engineers
Biological danger, risk of infections from epidemic diseases
Electrical shock risk: also 230 V can lead to death
High temperature
• Before operating, follow the instructions given at: " Signals and warnings" and
"Notice and biological danger".
• Address to engineers authorised by the manufacturer and use original spare parts
only. The manufacturer will not consider as under guarantee any appliance tampered
with by non-authorised engineers.
• Test suction by measuring capacity and head; a limited performance is caused
by something that must be removed: clogging caused by bacterial patinas, air loss
in dentist equipment, cracked or imploded tubes, aspiration stop due to siphons
in machine rooms or along the tubes; the fi rst can be eliminated by aspiring Puli-Jet
according to the instructions (it is sometimes necessary to repeat the same operation for
hours); in the second case, tubes and faulty parts must be replaced.
• When the aspiration unit is working, close the unit aspiration inlet and check the
operation of the compensation valve. (fi g. 7/A detail B).
• Take off the draining pipe (fig. 9 det. 6); aspirate plenty of water and check that
the centrifugal separator drains off with the correct pressure, verify that no water
comes out from exhausted air pipe.
• Check the electric switchboard. Check all the functions with various manoeuvres,
replace burnt pilot lights and worn relays. In case of equipment installed in parallel,
(fi g. 7/A-8-9 page 55-57-58)