• Notice
• Maintenance and cleaning
During surgical operations, foam build-up, caused by blood
and aspirated air, can reach probe level causing the unit to
stop; in this case we suggest the use of our solid anti-foa-
ming (directions for use are inside the package - Fig. B1). In
case some failure (clogging of cooling system, breakdown
etc.) should cause overheating of the motors - >120 °C for
Uni-Jet 75 suction unit and >90 °C for draining pump (8) -
a thermal device rated at a fixed temperature opens the
circuit and resets it automatically, when the temperature of
the windings is back to normal. Should this happen, identi-
fy and eliminate the cause.
Notice
Prior to starting any servicing operation on appliances
that have been used, clean with Puli-Jet plus new or
Eco-Jet 2 a few times as explained in the maintenance
section. Use disposable GLOVES (fig. B3), GOGGLES,
MASK and OVERALL.
Maintenance and cleaning
• Electrical shock risk: also 230 V can be lethal.
• Biological danger, risk of infections from epidemic
diseases.
• High temperature.
• Compulsory direction of flow and of rotation.
Together with liquids some solid particles may be su-
cked in, therefore it is necessary to have filters in order
to protect the motor and recover wanted particles.
γ
Aspi-Jet 6
is provided, like all other models, with a debris
filter 9 (Fig. C) which can be checked from outside the cabinet;
γ
Aspi-Jet 7
is provided with a filter (10) on the canister cover.
Filters must be cleaned every day. In order to remove
the filter 9, first turn the unit on and aspirate only air for
some seconds so that hoses and manifold dry out; di-
sconnect the power supply line, lift the terminals from their