DJO ARTROMOT-E2 Instrucciones De Uso página 84

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 138
11. Then put the enclosed polystyrene
block on the chair, observing the cut-
outs in the block.
12. Put the programming unit (15) in
the supplied box. Place the motion
element, the armrest and the power
cord in the respective cut-outs in the
polystyrene block of the package.
13. Then close the box.
Reverse the above steps to reassemble
the device after transport.
Warning!
m
Shock hazard –
Allow the ARTROMOT®-E2 to reach
room temperature before use. If the
device has been transported at temper-
atures below 0 °C (32°F), leave it to dry
at room temperature for about 2 hours,
until any condensation has disappeared.
6.3.2 ARTROMOT®-E2 compact
The following operating steps must
be completed before transporting the
ARTROMOT®-E2 compact:
1. Activate the "transport setting" function
in the menu (see also 5.3) and
start the ARTROMOT®-E2 compact.
2. Turn off the ARTROMOT®-E2 compact
power switch (19).
3. Remove the power cord (18) and dis-
connect the motion element (9) and the
programming unit (16).
4. Remove pin (26) on the underside and
take off the upper arm support.
5. Remove pin (5) and take off the swiv-
eling mechanism of the upper arm
support. Set the guide of the upper arm
support to the transport setting (paral-
lel with motor A).
6. Grasping the motion element at the
bracket (27) for adjustment of the posi-
tioning angle, open the locking screw (1)
and pull out the motion element.
7. Open the locking screw for adjustment
of the elbow angle adjustment (2) and
separate the oval tube from the motion
element.
8. Open locking screw (24) on the un-
derside of the stand, then remove the
column with the electronics box from
the star-shaped stand.
9. Disassemble the stand by loosening the
five legs with the special tool supplied
and remove them.
10. Unscrew the locking screw (7) from
motor B (12), remove the entire forearm
support including the slider (4) and
the hand support (13 + 14). Afterwards
screw the locking screw (7) back into
motor B.
11. Only use the original shipping box for
transporting the device. Ormed GmbH
& Co. KG cannot be held liable for
transport damage, if the device was not
shipped in its original shipping box.
12. Put the programming unit (15) in the
supplied box, using sufficient padding.
79
Tabla de contenido
loading

Este manual también es adecuado para:

Artromot-e2 compact

Tabla de contenido