12.2 Electromagnetic immunity
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-E2/-E2 compact is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure that the
ARTROMOT®-E2/-E2 compact unit is used in such an environment.
Immunity test
IEC 60601
test level
Home healthcare
Electrostatic
± 8 kV contact
discharge (ESD) to
± 15 kV air
IEC 61 000-4-2
Electrical fast
± 2 kV
transient/bursts to
100 kHz repetition
IEC 61 000-4-4
frequency
Surges to
± 1 kV
IEC 61 000-4-5
line-to-line
voltage
Voltage dips, short
0% U
interruptions and
at
voltage variations
0°, 45°, 90°, 135°,
on power supply
180°, 225°, 270°,
input lines to
315°
IEC 61000-4-11
0% U
and
70% U
cycles
Single phase at 0°
Voltage
0% U
interuptions to
cycles
IEC 61000-4-11
Power frequency
30 A/m
(50/60 Hz)
magnetic field to
IEC 61 000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
t
Compliance
level
Home healthcare
± 8 kV contact
± 15 kV air
± 2 kV
100 kHz repetition
frequency
± 1 kV
line-to-line
voltage
; 0,5 cycle
0% U
; 0,5 cycle
t
t
at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
; 1 cycle
0% U
; 1 cycle
t
t
and
; 25/30
70% U
t
t
cycles
Single phase at 0
; 250/300
0% U
; 250/300
t
t
cycles
30 A/m
268
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power should be that
of a typical commercial or
hospital environment.
Mains power should be that
of a typical commercial or
hospital environment.
Mains power should be that
of a typical commercial or
hospital environment. If the
user of the ARTROMOT®-E2/
-E2 compact unit requires
continued operation during
power mains interruptions,
it is recommended that the
; 25/30
ARTROMOT®-E2/-E2
compact unit be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristics of a typical
location in a typical commer-
cial or hospital environment.