– The programmed treatment para-
meters are deleted.
– All values stored on the chip card
are deleted.
– The carriage stops in the middle
position of the extension/flexion
and pronation/supination angles.
The "new patient" function (home
position) selects the following settings:
• extension:
• flexion:
• pronation:
• supination:
• pauses:
• timer: continuous operation
• speed:
• reverse on load motor A: 25
• reverse on load motor B: 25
• motor A:
• motor B:
• synchronized mode:
• total therapy time:
• special functions:
n Operating mode
synchronized/non-synchronized
Motors A and B can run in synchronized
or non-synchronized mode.
synchronized:
Motors A and B perform a synchronized
movement according to the elbow
joint's physiological motion pattern:
Starting at the middle position of the
set extension/flexion and pronation/
supination angles, the carriage first
moves simultaneously to the maximum
extension value and the maximum
pronation value. Afterwards it moves
simultaneously to the maximum flexion
and maximum supination values. After
reaching this position, the cycle restarts:
maximum extension and maximum
pronation.
During operation the synchronized
mode is indicated in the top left-hand
corner by the symbol
non-synchronized:
Both motors run independently of each
other within the selected ranges of
motion.
To enable the non-synchronized mode,
disable the synchronized mode.
During operation the non-synchronized
mode is indicated in the top left-hand
corner by the symbol
Default: synchronized mode enabled
m
25°
Patient hazard –
35°
The synchronized operation is recom-
mended as a general rule. However,
-5°
medical and/or therapeutic indications
may call for the non-synchronized
mode. In the non-synchronized mode,
0
particular caution and attention on
the part of the physician/therapist is
required to rule out any risk for the
100 %
patient.
n Service MENU
ON
For service purposes only, refer to Service
ON
Manual.
ON
Reminder:
0
You save the selected parameter values by
disabled
pressing the STOPkey.
5.4 Application and
5.4.1 Isolated extension/
1. Complete the mechanical settings and
2. Press the MENU key on the program-
3. Press the pronation (
4. Next set the desired range of motion
.
5. Set the flexion value (
74
Caution!
programming examples
flexion
adjust the carriage to the patient as
described in section 4.2.
ming unit to access level 1 (M1).
tion parameter key (
the + / - keys to set the carriage to
the appropriate rotation position to
be main tained during the isolated
extension/flexion movements.
for extension/flexion by first pressing
the extension parameter key (
then set the values with the + / - keys.
same way.
.
) or supina-
), the press
),
) in the