Precision
Precision studies using control solutions (intermediate precision) and
blood samples (repeatability) are shown blow:
Intermediate precision
Control
solutions
Mean
(mg/dL)
SD
(mg/dL)
CV (%)
Repeatability
Blood
samples
Mean
(mg/dL)
SD
(mg/dL)
CV (%)
FORA GTel β-Ketone Test Strips
Sample Size: 0.8 μL
Reaction Time: 10 seconds
System Measurement Range: 0.1 to 8.0 mmol/L
Hematocrit Range: 20% to 70%
Accuracy
FORA GTel system were tested in the hand of 120 lay users using
capillary blood samples, covering the rage between 0.04–6.77 mmol/L.
The results are compared to the laboratory method and are shown
below:
Level 1
Level 2
49.2
95.9
2.06
1.98
4.18%
2.07%
Level 1
Level 2
47.3
80.2
1.72
2.23
3.64%
2.78%
Level 3
Level 4
135.2
225.2
2.21
2.64
1.64%
1.17%
Level 3
Level 4
120.2
200.4
3.26
4.35
2.71%
2.17%
Level 5
315.5
4.03
1.28%
Level 5
352.5
9.19
2.61%
43