PERFORMANCE CHARACTERISTICS
FORA GTel Blood Glucose Test Strips
Sample Size: 0.5 μL
Reaction Time: 5 seconds
System Measurement Range: 10 to 600 mg/dL
Hematocrit Range: 20% to 70%
Accuracy
The FORA GTel meter result may vary slightly from your actual blood
glucose value. This may be due to slight differences in technique and
the natural variation in the test technology.
The chart below shows the results of a study where 350 typical users
used the FORA GTel meter to test their blood glucose level. For example,
in this study, the FORA GTel meter gave results within 15% of their true
blood glucose level 349 out of 350 times.
Difference range
between the true blood
glucose level and the
FORA GTel meter result.
The percent (and number)
of meter results that
match true blood glucose
level within x%
42
Within
Within
±5 %
±10 %
71.1%
94.9%
(249/
(332/
350)
350)
Within
Within
±15 %
±20 %
99.7%
100%
(349/
(350/
350)
350)