E Ensure nothing, such as a plaster, can impede the light passage, or else it
may result in inaccurate measure of SpO
F Excessive ambient light may affect the measure result. It includes fluores-
cent lamp, dual ruby light, infrared heater, direct sunlight and etc.
G Exquisite action of the subject or extreme electrosurgical interference may
also affect the accuracy.
H Testee can not use enamel or other makeup.
I Please clean and disinfect the device after operating according to the User
Manual (7.1).

6.3 Clinical restrictions

A As the measure is taken on the basis of arteriole pulse, substantial pulsat-
ing blood flow of subject is required. For a subject with weak pulse due to
shock, low ambient/body temperature, major bleeding, or use of vascular
contracting drug, the SpO
the measure will be more sensitive to interference.
B For those with a substantial amount of staining dilution drug (such as meth-
ylene blue, indigo green and acid indigo blue), or carbon monoxide he-
moglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and
some with icterus problem, the SpO
inaccurate.
C The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may
also be a major factor resulted in serious error of SpO
D As the SpO
2
oxia and toxic anoxia, some patients with serious anemia may also report
good SpO
measure.
2
7. MAINTENANCE, TRANSPORTATION
AND STORAGE

7.1 Cleaning and disinfecting

Using medical alcohol to disinfect the device, nature dry or clean it with clean
soft cloth.

7.2 Maintenance

A Please clean and disinfect the device before use according to the User
Manual (7.1).
B Please recharge the battery when the screen shows low power alarm infor-
mation.
C Recharge the battery soon after the over-discharge. The device should be
recharged every six months when it is not regular used. It can extend the
battery life following this guidance.