Press the Power button a second time to place the device in
operating mode� This second press helps prevent unwanted starts
or inadvertent changes in oxygen flow� The Introductory screen
momentarily appears, shown at left below, followed by the final
operating mode screen, shown at right�
Every time it starts up, the device starts producing oxygen to meet
the specified setting� It may take up to twenty minutes to reach the
specified purity levels for the setting� It is safe to begin breathing
from the simplyFLO device immediately�
Note: When it is turned on, the device will automatically start at the flow
rate setting used when the device was last turned off. As a precaution,
each time you start the device, verify the flow setting�
8�
To set the device's flow setting, press either the + or – button�
To increase the flow setting, press the + button� To decrease the flow
setting, press the – button�
The device can be set to �5, 1, 1�5, and 2 liters per minute (LPM) of
continuous flow�
Warning: It is very important to set your device to your prescribed level
of oxygen flow� Do not increase or decrease your flow rate from
your prescribed level until you first consult with your health care
professional�
9. Be sure oxygen is flowing through the cannula. If it is not, refer to the
Troubleshooting section of this manual�
D
escrIptIon
Push Power again to turn the device on�
Total number of operating hours for the device�
Software version in the device�
Indicates the allowable modes configuration
of the device� Note: This may not display on all
devices� This is normal�