Emc Information - Philips Respironics simplyFLO Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

simplyFLO User Manual

EMC Information

Guidance and Manufacturer's Declaration – Electromagnetic Emissions
simplyFLO is intended for use in the electromagnetic environment specified below� The user of simplyFLO should assure that it is
used in such an environment�
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
simplyFLO is intended for use in the electromagnetic environment specified below� The user of simplyFLO should assure that it
is used in such an environment�
Immunity Test
IEC 60601 Test Level
Electrostatic
±6 kV Contact
Discharge (ESD)
±8 kV Air
IEC 61000-4-2
Electrical Fast
±2 kV for Power Supply
Transient/Burst
Lines
IEC 61000-4-4
±1 kV for Input/Output
Lines
Surge
±1 kV Line to Line
IEC 61000-4-5
±2 kV Line to Ground
Voltage Dips, Short
<5% U
Interruptions and
for 0�5 Cycle
Voltage Variations on
Power Supply Input
40% U
Lines
for 5 Cycles
IEC 61000-4-11
70% U
for 25 Cycles
<5% U
for 5 Seconds
Power Frequency
3 A/m
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
Note: U
is the A�C� mains voltage prior to application of the test level�
T
20
Compliance
Group 1
Class B
Class A
Complies
Compliance Level
±6 kV Contact
±8 kV Air
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
±1 kV Line to Line
±2 kV Line to Ground
(>95% Dip in U
)
<5 % U
T
T
for 0�5 Cycle
(60% Dip in U
)
40 % U
T
T
5 Cycles
(30% Dip in U
)
70 % U
T
T
25 Cycles
(>95% Dip in U
)
<5 % U
T
T
for 5 Seconds
3 A/m
Electromagnetic Environment - Guidance
The Device uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment�
The Device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes�
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%�
Mains power quality should be that of a typical
home or hospital environment�
Mains power quality should be that of a typical
home or hospital environment�
(>95% Dip in U
)
Mains power quality should be that of a typical
T
T
home or hospital environment� If the user of the
Device required continued operation during
(60% Dip in U
) for
power mains interruptions, it is recommended
T
T
that the Device be powered from an
uninterruptible power supply or battery�
(30% Dip in U
) for
T
T
(>95% Dip in U
)
T
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical home or hospital environment�
Tabla de contenido

Solución de problemas

loading

Tabla de contenido